Ermed Georgia

ARCHITECT ci4100

SMALLER VOLUME
HIGH PRODUCTIVITY

The ARCHITECT ci4100 combines ARCHITECT c4000 and ARCHITECT i1000SR instruments − offering lower volume laboratories the enhanced efficiency and productivity employed by larger volume laboratories.

The ARCHITECT ci4100 offers a maximum throughput of up to 900 tests per hour, including 800 clinical chemistry and 100 immunoassay tests. Featuring a load-up capacity of 180 samples with 35 priority positions, the ARCHITECT ci4100 has up to 115 refrigerated reagent positions plus Integrated Chip Technology (Na+, K+ and Cl-).

DESCRIPTION
Methods See c4000 and i1000SR specifications
Maximum Throughput Up to 900 tests/hour
Sample Types See c4000 and i1000SR specifications
Sample Tubes Height: 72-102 mm
Diameter 9.6-16.1 mm
Sample Cup Yes (50 µL dead volume)
Sample Capacity 180
Sample Barcode Types Code 39, Codabar, Interleaved 2 of 5, Code 128
Sample Result Storage 50,000
Sample Volume See c4000 and i1000SR specifications
Automatic Dilution Yes
Sample Probe Carryover ≤0.1 parts per million
Reagent Capacity Up to 115 refrigerated positions plus patented ISE (Na+, K+ and Cl-)
Reagent Type See c4000 and i1000SR specifications
Reagent Onboard Stability See c4000 and i1000SR specifications
Calibration Frequency See c4000 and i1000SR specifications
Sample, Clot and Bubble Detection  Yes
Reagent Pressure Monitoring See c4000 and i1000SR specifications
Sample Interference Measurement Yes: Hemolysis, Icterus, and Lipemia
System Control Center  1 SCC, with color touchscreen monitor, keyboard and mouse
Onboard Maintenance Records Yes
Online Error Code Help Yes
Host Interface Bidirectional, serial RS-232 interface, host query option available
Remote Diagnostics AbbottLink
Dimensions
(H x W x D)
49” x 111” x 36”
125.1 x 281.2 x 90.7 cm
Weight 1,677 lbs
760.7 kg
Electrical Requirements See c4000 and i1000SR specifications
Water Requirements See c4000 and i1000SR specifications
Heat Output* See c4000 and i1000SR specifications
Sample Loading RSH
*Values provided represent the typical output in “Running” mode for the system processing module and sample handler. A maximum value of 389 BTU/hr was obtained for the System Control Center (SCC) using the ARCHITECT i2000SR as a representative system. Values obtained during an internal study. Data is on file with Abbott Diagnostics.
Note: Some of these specifications are based on the combination of the ARCHITECT c4000 and the ARCHITECT i1000SR. For accurate and complete specifications, please refer to the individual operations manual.